Limitations and accuracy
Limitations
There may be some limitations in
the use of the pressure bomb.
- If the stems of plants are very
weak, vessels are obliterated by the pressure of the sealing sleeve into
the specimen holder. The reading is then in accurate because the sap can
not flow easil past the constriction to the cut end.
- Even though this device is supplemented
with different diameters of rubber sleeves, a good seal may not be possible
if plants are very small.
- If the pressure bomb readings are
to be used a to schedule irrigation, it is necessary to run a large number
of samples in order to compensate for variability of the plant status.
- The use of the pressure bomb may
be limited when the internal water status of plants is under equilibrium.
Data is highly affected by pre-dawn conditions.Measurements should be taken
when internal water status is in some way stabilized.
Accuracy
When the pressure technique is carefully executed,
it will give the sap pressure that existed at the time the sample was cut.
The pressure at which sap comes through the end of the branch is equivalent
to the negative pressure that existed in plant when the sample was cut.
Therefore, pressure bomb accuracy is high. However, some factors
may influence accuracy such as:
- Evaporation loss from the twig must to be avoided.
Keep all the water inside the plant tissue before the reading is taken
to avoid turgor loss inside the leaves, which in turn may give a wrong
reading.
- When sampling, sample size should match the pressure
vessel size to avoid a new sample cutting or a trimming.
- In order to increase the level of accuracy it
is necessary to carry out a large number of samples. In other words, the
higher the number of samples, the higher the accuracy.
- It is important to consider the period of time
from the sample cutting until the final reading of the pressure bomb. Consistent
durations of time per sample results in higher levels of accuracy.
- If ice is used to keep plant tissues cool enough
to avoid moisture loss, the ice must not touch samples because may cause
membrane damages, which may produce inaccurate readings.
- Also, twigs exposure time during the time of
the day may affect greatly the accuracy because there is more variation
in leaf water potential during morning than at midday.
- It is essential to use healthy samples, free
of diseases, insects, mechanical and other kind of damages.
Go back to the top
Back to the
main document