Gas Ignition

| A Fuel Pressure Gauge | G Oxidant
Pressure Regulator |
| B Oxidant Pressure Gauge | H Fuel
Flow Needle Valve |
| C Fuel Toggle Valve | I Oxidant
Selector Valve |
| D Fuel and Oxidant Flowmeters | J Oxidant
Flow Needle Valve |
| E Ignite Button | K Power
ON/OFF Switch |
F Fuel Pressure Regulator |
1. Turn the POWER switch of the gas
control to the ON position.
2. Turn on the vent hood to remove fumes and vapors from the flame.
3. Switch the OXIDANT SELECTOR VALVE to air. The air supply for the AA in room 263
comes from a central source and an oil-filter is used to maintain the air free of oil and
dust (Please check Special Laboratory Requirements section in Instruction
Manual).
4. Switch the FUEL VALVE to the on position (For most of the analyses acetylene is the
fuel of choice for Perkin-Elmer AA, however, if some other fuel is required please check
Special Laboratory Requirements section in The Instructions Manual).
5.Once the fuel and oxidant flow meter are set at 30 and 55 (arbitrary units)
respectively, ignite the burner by pressing the IGNITE button. Ignition will occur
only when the oxidant selector valve is on the air position and the safety interlock pin
on the burner head hold-down cable is connected.